LCMS - Psychotropic Medication 1, tricyclische antidepressiva TDM

Psychotropic Medication 1 (TCA) Reagent Set

SKU: 4050 P PS1

Voordelen van de Psychotropic Medication 1 LC-MS kit

  • Gebalanceerd panel dat optimaal is voor de meest gevraagde TCA-componenten
  • Eenvoudige en kolomvriendelijke monstervoorbereiding
  • Geen SPE of LLE meer nodig
  • Interne standaard voor alle hoofdcomponenten (alleen niet voor Z-10-OH- en E-10-OH-nortriptyline)
  • 25 µl monster nodig (t.o.v. 500 µl bij SPE-HPLC-UV). Minder volume is waarschijnlijk haalbaar.

Components

  • Amitriptyline
  • Nortriptyline
  • Z-10-OH-nortryptiline
  • E-10-OH-nortyptiline
  • Clozapine
  • Norclozapine
  • Clomipramine
  • Norclomipramine
  • Imipramine
  • Desipramine
  • Zuclopentixol

In het kort

  • Matrix: Serum
  • Technique: LC-MS/MS
  • Internal Standard: Amitriptyline-D3, Clozapine D4, Clomipramine D3, Desipramine D4, Imipramine D4, Nortriptyline D3 (also used for E-10-OH-Nortriptyline & Z-10-OH-Nortriptyline) & Zuclopenthixol-D4

Beoogd gebruik



De Diagnotix Psychotropic Medication 1 LC-MS Reagent Kit is ontworpen om de concentratie van een of meer van de genoemde analyten en metabolieten in serum of plasma te bepalen als basis voor verdere diagnostiek.



Technische voordelen


  1. Gebalanceerd panel: De meest voorkomende verbindingen en hun gemeenschappelijke metabolieten worden gekozen om ervoor te zorgen dat het panel niet te groot is, maar toch voldoet aan de uitdagingen van routinematig gebruik. Hoe groter het panel, hoe meer kosten er zijn om alle analyten en hun isotopen op te nemen, maar ook hoe meer afwegingen er moeten worden gemaakt voor chromatografische optimalisatie.
  2. Uniek panel: Gebasseerd op de SKML panels PS I en II. Op verzoek van Nederlandse specialisten hebben we E- en Z-OH nortriptyline aan het panel toegevoegd.
  3. Kolom- en LC-MS-vriendelijk: Er wordt een 40-voudige deproteïnisatieoplossing gebruikt, wat zorgt voor extreme monsterreiniging. Ook is voor de meeste systemen slechts een klein injectievolume vereist.
  4. Kosteneffectieve en snel automatiseerbare monsterbereiding: Er is geen SPE meer nodig. SPE voor HPLC-UV-analyses worden gebruikt om de levensduur van de kolom te verlengen en UV-interferentie van mogelijke interfererende drugsmisbruik te verminderen. Zoals vermeld in 3. wordt het monster 40 keer verdund en wordt in principe alle proteïne verwijderd. Bovendien zorgt het gebruik van LC-MS ervoor dat geen co-eluerende moleculen interfereren met uw verbinding van belang, omdat ze niet dezelfde massatransities delen.


Waarom overstappen van HPLC naar LC-MS?


Door over te stappen op LC-MS bespaart u direct op het gebruik van SPE, langere monstervoorbereiding en dus arbeidskosten. Bovendien zal de runtijd hoogstwaarschijnlijk een stuk korter zijn en hoeft u zich geen zorgen te maken over interfererende andere drugs, zolang ze niet dezelfde massatransities delen.

En omdat basislijnscheidingen niet langer nodig zijn en de mobiele fasen geen buffers met een hoge concentratie hoeven te bevatten, profiteert u van een veel langere levensduur van de kolom.


Vooral interessant voor laboratoria die:

I. Regelmatig (grote) batches TCA-patiëntmonsters hebben en een betrouwbare automatiseerbare workflow hebben.

II. Hebt u al een LC-MS-systeem, maar voert u nog steeds TDM- of TCA-analyses uit op HPLC-UV-machines. Het elimineren van het gebruik van uw HPLC-UV-systeem verbetert onmiddellijk de operationele efficiëntie en verlaagt de totale onderhoudskosten.


Neem contact met ons op voor meer informatie

Wilt u meer weten over onze Psychotropic Medication 1-oplossing voor Tryciclic-antidepressiva? Neem contact met ons op via het onderstaande contactformulier. We horen graag van u!

Contactformulier

Kit wordt geleverd inclusief:

  • 4051 CAL P PS1, Calibrator Set (Calibrator 1 – 6), 6 x 2 x 500 µl
  • 4059 P PS1, Deproteinization Solution + Internal Standard, 3 x 100 ml
  • 4064 P PS1, Mobile Phase I, 1 x 500 ml
  • 4065 P PS1, Mobile Phase II, 1 x 500 ml
  • 4066 P PS1, Autosampler Washing Solution, 1 x 1000 ml

Klinische relevantie


Tricyclic antidepressants (amitriptyline, nortriptyline, clomipramine, imipramine and desipramine) and antipsychotics (clozapine). Therapeutic Drug Monitoring (TDM) is based on the assumption that there is a relationship between the blood concentration and clinical effect (therapeutic improvement and adverse effects). It also assumes there is a concentration range of the drug which is characterized by maximal effectiveness and maximal safety, the “therapeutic window”. The Diagnotix kit for measuring tricyclic antidepressants (TCAs) and antipsychotics (PS1) includes the TCAs amitriptyline, nortriptyline, clomipramine, imipramine and desipramine (including the drugs’ metabolites) and the antipsychotic clozapine (including the metabolite norclozapine).The TCAs amitriptyline, nortriptyline, clomipramine, imipramine and desipramine are used for the treatment of several types of depression, obsessive compulsive disorder, neuropathic pain, nocturnal enuresis, and for the prophylactic treatment of chronic tension type headache and migraine. TCAs have a pronounced inter-individual pharmacokinetic variability and a narrow therapeutic window. Studies on the relation between blood concentration and clinical improvement have supported the relation for the TCAs. Systematic reviews and meta-analyses led to convincing evidence of a significant relationship between clinical outcomes and plasma concentrations for the TCAs which are associated with a high probability of response.

Chromatogrammen

Technische data

Toepasbaar op:
Limit of Quantification; Amitriptyline / Clomipramine / Clozapine / Desipramine / Imipramine (µg/l) 3.66 / 6.12 / 11.05 / 4.27 / 4.97
Norclomipramine / Norclozapine / Nortriptyline / E-10-OH-Nortriptyline / Z-10-OH-Nortriptyline / Zuclopenthixol (µg/l) 4.31 / 6.46 / 5.65 / 15.49 / 4.10 / 2.63
Linearity; Amitriptyline / Clomipramine / Clozapine / Desipramine / Imipramine (µg/l) 1800 / 1400 / 2000 / 1230 / 2000
Norclomipramine / Norclozapine / Nortriptyline / E-10-OH-Nortriptyline / Z-10-OH-Nortriptyline / Zuclopenthixol (µg/l) 1175 / 2650 / 1250 / 1050 / 288 / 290
Accuracy Passes (<15% CV)
Simple Precision; Amitriptyline / Clomipramine / Clozapine / Desipramine / Imipramine (µg/l) < 3.7 / < 2.0 / < 2.4 / < 2.8 / < 4.1
Norclomipramine / Norclozapine / Nortriptyline / E-10-OH-Nortriptyline / Z-10-OH-Nortriptyline / Zuclopenthixol (µg/l) < 2.3 / < 3.3 / < 6.6 / < 4.0 / < 5.3 / < 10.9
Complex Precision; Amitriptyline / Clomipramine / Clozapine / Desipramine / Imipramine (µg/l) < 8.0 / < 6.4 / < 4.9 / < 5.5 / < 7.1
Norclomipramine / Norclozapine / Nortriptyline / E-10-OH-Nortriptyline / Z-10-OH-Nortriptyline / Zuclopenthixol (µg/l) < 6.8 / < 5.0 / < 10.6 / < 8.1 / < 8.1 / < 13.9
Ionization Electrospray
MS/MS mode Positive
Sample Preparation 25 µl sample → 1 ml deprot + IS → centrifuge → inject
Aanbevolen Column Acquity UPLC HSS T3 1,8 µm 2.1x100 mm
Analysis Time 7 min
Flow Rate 0.4 ml/min
Injection Volume 2-20 µl
Validation Report Available Yes

Please Note:Information as presented freely on this website is not sufficient for proper operation of these medical devices. Please refer to the applicable instructions for use and all accompanying information such as warning notices. All necessary documentation can be found in the download centre section of this website.For other LC-MS/MS systems than listed above, please contact us to discuss the possibilities.

Geinteresseerd in onze producten of diensten?
Neem contact op
of bel ons op +31 (0)596 20 10 62

Meer informatie?

* Verplichte velden

Neem voor meer informatie over dit product contact met ons op door middel van het contactformulier of onderstaande contactgegevens.

Contactgegevens

De Plassen 4
9902 SE Appingedam
Nederland

info@diagnotix.com +31 (0)596 20 10 62

Gerelateerde producten