The Cortisol & Cortisone reagents set from diagnotix • •• can determine the concentrations of both cortisol and cortisone in urine or saliva.

Advantages of the Diagnotix Cortisol & Cortisone Urine LC-MS D&S method:

  • 6-point calibrator based on Human urine like matrix
  • Internal standard with deuterium labeled cortisol and cortisone
  • Our extremely simple sample preparation process helps you save both time and costs, making your workflow more efficient and cost-effective
  • Easy to automate

Intended use



The Diagnotix Cortisol & Cortisone LC-MS D&S Kit is specifically designed to measure cortisol and cortisone concentrations in patient samples, providing a reliable foundation for further diagnostic evaluations.



Technical benefits


  1. Simplified sample preparation: A simple dilution is enough for fast and reliable analyses.
  2. Savings: Saving preparation time and eliminate the need of expensive SPE-tubes and workflow.


Especially interesting for:

  • Laboratories processing relatively small sample volumes on a biweekly or monthly basis.
  • Laboratories with sufficient LC-MS knowledge that understands the importance of optimized and frequent LC-MS cleanup schedules.
  • Facilities operating multiple LC-MS systems, allowing for dedicated systems for specific methodologies.


Contact Us for More Information


Interested in learning more about our Cortisol&Cortisone D&S analysis? Contact us via the contact form below. We looking forward to hear from you!

Contact form

Included in kit (300 assays):

  • 1042 CAL M CORT, Calibrator Set (Calibrator 1 – 6), 6 x 2 x 1 ml
  • 1049 M CORT, Internal Standard, 3 x 6 ml
  • 1190 M CODS, Dilution Solution 3 x 100 ml
  • 1191 M CODS, Mobile Phase I, 1 x 500 ml
  • 1192 M CODS, Mobile Phase II, 1 x 500 ml
  • 1196 M CODST, Autosampler Washing Solution, 1 x 1000 ml

Clinical relevance

The ratio between cortisol and cortisone is of clinical importance with the diagnosis of Cushing syndrome. This even indicates apparent mineralocorticoid excess, a disease where the 11β-hydroxysteroid dehydrogenase-activity is weakened significantly or is even absent.

Chromatograms

Technical data

Number of samples kit 300
Matrix Urine
Sample Preparation • Centrifuge your sample
• Pipette 100 µl into vial
• Add 900 µl Dilution Solution and mix well
• Transfer into LC-MS vial or 96-wellplate
• Ready to analyze
Limit of Detection (LOB) • Cortisol: 0 nmol/l
• Cortisone: 0.218 nmol/l
Linearity • Cortisol: 1200 nmol/l
• Cortison: 3100 nmol/l
Repeatability (CV%) • Cortisol control low: 2.0
• Cortisol control high: 1.0
• Cortisol patient: 1.4
• Cortisone control low: 3.0
• Cortisone control high: 2.5
• Cortison patient: 0.9
Reproducibility (CV%) • Cortisol control low: 3.0
• Cortisol control high: 2.5
• Cortisol patient: 3.2
• Cortisone control low: 2.9
• Cortisone control high: 2.6
• Cortisone patient: 3.4
Injection volume 5 - 20 µl
Ionisation ESI Postive
Examples of LC-MS systems • Waters TQS-Micro or more sensitive
• Shimadzu 8050 of more sensitive
• Sciex 5500 of more senstive

Please Note:



Information as presented freely on this website is not sufficient for proper operation of these medical devices. Please refer to the applicable instructions for use and all accompanying information such as warning notices. All necessary documentation can be found in the download centre section of this website.For other LC-MS/MS systems than listed above, please contact us to discuss the possibilities.

Interested in our products or services?
Contact us
or call us on +31 (0)596 20 10 62

More info?

* Mandatory fields

Contact us for more information about this product using the contact form or the contact information below.

Contact details

De Plassen 4
9902 SE Appingedam
The Netherlands

info@diagnotix.com +31 (0)596 20 10 62

Related products